2-162 Chapter 2
System Verification and Performance Tests
Agilent 8753ET Performance Test Records
Agilent Technologies Company
Model 8753ET
Serial Number
Report Number
Option(s) Date
1. Reflection Test Port Output Frequency Range and Accuracy
Note: If your analyzer does not have Option 006, write “N/A” for entries above 3 GHz.
CW Frequency
(MHz)
Min. (MHz) Measured Value
(MHz)
Max.
(MHz)
Measurement
Uncertainty
(MHz)
0.3 0.299 997 0.300 003 ± 0.000 000 520
5.0 4.999 950 5.000 050 ± 0.000 008 610
16.0 15.999 840 16.000 160 ± 0.000 028 220
31.0 30.999 690 31.000 310 ± 0.000 053 730
60.999 999 60.999 389 61.000 609 ± 0.000 104 800
121.0 120.998 790 121.001 210 ± 0.000 206 800
180.0 179.998 200 180.001 800 ± 0.000 307 200
310.0 309.996 900 310.003 100 ± 0.000 528 300
700.0 699.993 000 700.007 000 ± 0.001 191 700
1 300.0 1 299.987 1 300.013 ± 0.002 212 300
2 000.0 1 999.980 2 000.020 ± 0.003 403 000
3 000.0 2 999.970 3 000.030 ± 0.005 104 000
4 000.0 3 999.960 4 000.040 ± 0.006 805 000
5 000.0 4 999.950 5 000.050 ± 0.008 506 000
6 000.0 5 999.940 6 000.060 ± 0.010 207 000

Table of Contents

Other manuals for Agilent Technologies 8753ET

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8753ET and is the answer not in the manual?

Agilent Technologies 8753ET Specifications

General IconGeneral
BrandAgilent Technologies
Model8753ET
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals