3-46 Chapter 3
Adjustments and Correction Constants
Frequency Accuracy Adjustment
Figure 3-19 High Stability Frequency Adjustment Location
11.Insert a narrow screwdriver and adjust the high-stability frequency reference
potentiometer for a CW frequency measurement within specification.
In Case of Difficulty
Replace the A26 assembly if you cannot adjust the CW frequency within specification.