3-58 Chapter 3
Adjustments and Correction Constants
Sequences for Mechanical Adjustments
Sequences for the Fractional-N Avoidance and FM Sideband Adjustment
— Sequence APIADJ sets up the fractional-N API spur adjustments. —
TITLE
SP 2.5K
PERIPHERAL HPIB ADDR
18 x1
TITLE TO PERIPHERAL
WAIT x
0 x1
TITLE
AT 0DB
TITLE TO PERIPHERAL
WAIT x
0 x1
TITLE
RB 100HZ
TITLE TO PERIPHERAL
WAIT x
0 x1
TITLE
CF 676.145105MZ
TITLE TO PERIPHERAL
WAIT x
0 x1
CW FREQ
676.045105M/u
TITLE
ADJ A13 100KHZ
SEQUENCE
PAUSE
TITLE
CF 676.051105MZ
TITLE TO PERIPHERAL
WAIT x
0 x1
CW FREQ
676.048105M/u
TITLE
ADJ A13 API1
SEQUENCE
PAUSE