Chapter 2 2-37
System Verification and Performance Tests
Agilent 8753ES System Verification and Performance Tests
Figure 2-18 Setup for Power Meter Calibration on Test Port 1
3. Press .
4. Only for Analyzers with Option 006:
Press .
5. Press .
6. Press and until the analyzer shows the correct
power meter model. (Use the 438A/437 selection if the power meter is an HP/Agilent
E4419B or an E4418B.)
7. Press . The default power meter GPIB address is 13. Make
sure it is the same as your power meter GPIB address. Otherwise, use the analyzer
front panel keypad to enter the correct GPIB address for your power meter.
8. Press .
Preset Avg
IF BW
3
k/m
Start 300 k/m
Stop 3 G/n
Local
SYSTEM CONTROLLER
SET ADDRESSES
POWER MTR
ADDRESS: P MTR/GPIB
Sweep Setup
NUMBER of POINTS
51 x1