EasyManua.ls Logo

Agilent Technologies 8890 - Page 242

Agilent Technologies 8890
278 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
17 Chromatographic Checkout
242 Operation Manual
8 If using a data system, prepare the data system to perform one run using the loaded
checkout method. Make sure that the data system will output a chromatogram.
If not using a data system, create a one sample sequence using the GC Browser
Interface.
9 Start the run.
If performing an injection using an autosampler, start the run using the data system, or
press .
If performing a manual injection (with or without a data system):
a Select to prepare the inlet for splitless injection.
b When the GC becomes ready, inject 1 µL of the checkout sample and select on the
GC.
c The following chromatogram shows typical results for a new detector with new
consumable parts installed and nitrogen makeup gas.
ALS settings (if installed)
Sample washes 2
Sample pumps 6
Sample wash volume 8 (maximum)
Injection volume 1 µL
Syringe size 10 µL
Solvent A pre washes 2
Solvent A post washes 2
Solvent A wash volume 8
Solvent B pre washes 0
Solvent B post washes 0
Solvent B wash volume 0
Injection mode (7693A) Normal
Airgap Volume (7693A) 0.20
Viscosity delay 0
Inject Dispense Speed (7693A) 6000
PreInjection dwell 0
PostInjection dwell 0
Manual injection
Injection volume 1 µL
Data system
Data rate 5 Hz
Table 42 FID Checkout Conditions

Table of Contents

Other manuals for Agilent Technologies 8890

Related product manuals