CAPINTEC, INC. CAPTUS
®
600
Press
F1 [SKIP] to skip the measurement (i.e. perform it later if possible or choose not to
perform it).
Figure 8-18 will appear.
14:32
THYROID UPTAKE
MEAS
PRE-DOSE MEASUREMENT: SKIP
OK
Figure 8-18
Press
F4 [OK] to accept the displayed choice. If OK is pressed while “SKIP” is displayed,
you will be asked to verify that you want to skip the Pre-Dose Measurement.
Press
BSP to exit the screen without making a choice and ending the current procedure.
If Pre-Dose Measurement is chosen, the Measure Neck Screen (
Figure 8-19) will appear.
14:32
THYROID UPTAKE
DASILVA, AND
START
EXIT
PROTOCOL: 1 CAPSULE
MEASURE
NECK
AT: > 25 CM
COUNTING REPETITION # 1 / 2
Figure 8-19
If the Thyroid Uptake method is “ONE CAPSULE” or “TWO CAPSULES”, the displayed “AT:”
distance of 25cm is the default value of the Probe’s position away from the patient's neck. If
this is not the value to be used for the Thyroid Uptake study, input the correct value. This
value will be used throughout the study and cannot be changed.
If the Thyroid Uptake method is “STANDARD CAPSULE”, the displayed “AT:” distance is the
distance chosen when the Standard Capsule was first measured.
8-10 THYROID UPTAKE GENERAL INSTRUCTIONS November 07