CAPINTEC, INC. CRC
®
-25R
Slip Printer
• If the RELEASE light (on printer) is not lit, press RELEASE (on printer).
• Place Ticket in printer, face up.
• Slide the ticket under the mechanism until the PAPER OUT light is off.
• Input the User ID if desired.
• Press ENTER (on the readout).
Note: Tickets cannot be printed using an Inkjet printer.
MOLY ASSAY
The CRC
®
-25R steps the user through a Moly Assay procedure. The assay may be
performed in the CapMac or the Canister.
If one method (CAPMAC or CANISTER) is to be used all the time, choose the method in
Setup, Other, Moly Setup. (Reference CHAPTER 6: CHAMBER INITIALIZATION, SECTION:
MOLY ASSAY SETUP)
To perform a Moly Assay, press TEST from Figure 10-1 Measurement Screen.
Select MOLY ASSAY.
If the assay method has not been uniquely specified, Figure 10-9 Moly Assay Method
Selection screen will appear. The method must chosen first.
1. CAPMAC for
Mallinckrodt Gen
2. CAPMAC for
Bristol Myers Gen
3. Capintec Canister
Figure 10-9 Moly Assay Method Selection
Select the method to be used for the current test only.
If only one method (CAPMAC or CANISTER) has been chosen in Moly Setup, Figure 10-9
Moly Assay Method Selection will be skipped and Figure 10-10 Moly Assay Background
Choice will appear (example shown is for CAPMAC).
10 - 10 MEASUREMENT PROCEDURES March 15