EasyManuals Logo

Capintec CRC-25R User Manual

Capintec CRC-25R
228 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #138 background imageLoading...
Page #138 background image
CAPINTEC, INC CRC
®
-25R
Place the Bottom of Strip in the Chamber and press any key (except HOME) to perform the
measurement.
Figure 11-34 QC Test Measurement Screen will re-appear.
Press ENTER to accept the measurement.
Figure 11-36 1 Strip Results screen will appear.
QC TEST
Single Strip
Top /(Top+Bottom)
xx.x%
Bottom/(Top+Bottom)
xx.x%
Any Key to Continue
Figure 11-36 1 Strip Results
Press any key (except HOME) to continue. If a printer is attached to the system, a QC Test
1 Strip report will print. The report will show the measured values and results.
The prompt to print again will appear. To print the test results again, press YES. Press NO to
continue.
Figure 11-37 Perform Another Test Screen will appear.
Do Another
Test? Y or N
Figure 11-37 Perform Another Test Screen
To perform another test, press YES.
Figure 11-33 1 Strip Measurement Message – Top will
re-appear.
If another test is not desired, press NO to exit.
Figure 11-1 Tests Menu will appear.
Two Strips Test
The Two Strips Test consists of 4 measurements: the top and then the bottom of strip A, and
the top and then the bottom of strip B.
Select 2 STRIPS from
Figure 11-32 QC Test Selection Menu. Figure 11-38 2 Strips
Measurement Message – Top A
will appear.
11 - 16 ENHANCED TESTS August 07

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Capintec CRC-25R and is the answer not in the manual?

Capintec CRC-25R Specifications

General IconGeneral
BrandCapintec
ModelCRC-25R
CategoryLaboratory Equipment
LanguageEnglish