CAPINTEC, INC. CRC
®
-25R
Data Check
The next part of the Daily Test is a check of the built-in nuclide data. Figure 8-14 Data
Check Results screen will appear.
DATA CHECK
OK
Any Key to Continue
Figure 8-14 Data Check Results
If this test fails, turn the power off and then back on. This will reload the program and data
into memory. Repeat the test. If the test continues to fail, contact Capintec’s only
Authorized
Service Center at 1-800-227-6832.
Press ENTER to accept the result and continue the Daily Test.
Accuracy and Constancy Test in Daily Test
The next part of the Daily Test is the Accuracy Test. The CRC
®
-25R will step through
performing the Accuracy Test for all sources that have been designated as a Daily Source
(reference CHAPTER 6: CHAMBER INITIALIZATION, SECTION: TEST SOURCES).
Note: If no sources were designated as Daily Test sources, then the message “NO DAILY
SOURCE DATA – Any Key to Continue” will appear and the test will be completed.
The Accuracy and Constancy tests will not be performed.
In the example below, Co57 is the Daily Source and the Constancy Source.
MEASURE Co57
S/N:12345
Any Key to Continue
Figure 8-15 Accuracy Test in Daily Test Instruction Screen
Place the requested Test Source into the Chamber and press any key (except HOME) to
continue with the Accuracy Test. Figure 8-16 Accuracy Test Result Screen will appear and
is shown below along with a descript
i
on of each section.
October 13 BACKGROUND AND TESTS 8 - 7