CAPINTEC, INC CRC
®
-25R
The prompt to print again will appear. To print the test results again, press YES. Press NO to
continue.
Figure 11-44 Perform Another Test Screen will appear.
Do Another
Test? Y or N
Figure 11-44 Perform Another Test Screen
To perform another test, press YES.
Figure 11-38 2 Strips Measurement Message – Top A
will re-appear.
If another test is not desired, press NO to exit.
Figure 11-1 Tests Menu will appear.
HMPAO Test
The Tc99m HMPAO Test consists of 6 measurements: the top and then the bottom of strip
A, the top and then the bottom of strip B, and the top and then the bottom of strip C.
Select HMPAO from
Figure 11-32 QC Test Selection Menu. Figure 11-45 HMPAO
Measurement Message – Top A
screen will appear.
Place TOP of
Strip A in Chamber
Any Key to Continue
Figure 11-45 HMPAO Measurement Message – Top A
Place the Top of Strip A in the Chamber and press any key (except HOME) to perform the
measurement.
Figure 11-46 QC Test Measurement Screen will appear.
August 07 ENHANCED TESTS 11 - 19