CAPINTEC, INC. CRC
®
-25R
Moly Assay Limit
The Moly Assay default limit is set to 0.150 µCi/mCi (or 0.150 MBq/GBq). The user may
change this value.
Select MO/TC LIMIT from Figure 6-19 Moly Setup Menu. The current limit is displayed:
Mo/Tc LIMIT:
0.150 µCi/mCi
OK? Y or N
Figure 6-22 Mo/Tc Limit Verification Screen
If the displayed value is acceptable, press YES to keep the value and return to Figure 6-19
Moly Setup Menu.
If the displayed value is not acceptable, press NO. Figure 6-23 Mo/Tc Limit Entry Screen will
appear. The value is input in µCi/mCi (or MBq/GBq if in Becquerel mode):
INPUT
Mo/Tc Limit
in µCi/mCi
_
Figure 6-23 Mo/Tc Limit Entry Screen
Input the new value and press ENTER. Figure 6-22 Mo/Tc Limit Verification Screen will re-
appear displaying the new value.
Note: The minimum value that can be input is 0.001 µCi/mCi (0.001 MBq/GBq). The
maximum value that can be input is 9.99 µCi/mCi (9.99 MBq/GBq).
If the displayed value is acceptable, press YES to keep the value and return to Figure 6-19
Moly Setup Menu.
March 15 CHAMBER INITIALIZATION 6 - 13