Standard Test
The Standard Test measures the activity of Tc99m over a period of time. When sufficient
measurements have been made, a least-squares fit is performed and the deviation of the
results from this fit are reported.
First Measurement
Figure 11-14 Standard Linearity Test First Measurement Prompt will appear for the
first measurement.
Test # 1
First Measurement
Any Key to Continue
Figure 11-14 Standard Linearity Test First Measurement Prompt
Place the linearity source in the Chamber and press any key (except HOME) to
perform the measurement.
Figure 11-15 Standard Test Measurement Screen will
appear.
Tc99m
6.27mCi
ENTER to Accept
Figure 11-15 Standard Test Measurement Screen
Press ENTER to accept the measurement.
Figure 11-16 Standard Linearity Test First
Measurement Verification
will appear.
August 07 ENHANCED TESTS 11 - 7