CAPINTEC, INC CRC
®
-25R
Press ENTER to accept the measurement. Figure 11-52 HMPAO Results Screen will
appear.
QC TEST HMPAO
Free Tc99m
xx.x%
Hydrolized/Reduced
xx.x%
Tc99m HMPAO
xx.x%
An
y
Ke
y
to Continue
Figure 11-52 HMPAO Results Screen
Press any key (except HOME) to continue. If a printer is attached to the system, a QC Test
HMPAO report will print. The report will show the measured values and results.
The prompt to print again will appear. To print the test results again, press YES. Press NO to
continue.
Figure 11-53 Perform Another Test Screen will appear.
Do Another
Test? Y or N
Figure 11-53 Perform Another Test Screen
To perform another test, press YES.
Figure 11-45 HMPAO Measurement Message – Top A
will re-appear.
If another test is not desired, press NO to exit.
Figure 11-1 Tests Menu will appear.
MAG3 Test
The Tc99m MAG3 Test consists of 3 measurements: Fraction #1, Fraction #2 and the
Cartridge.
Select MAG3 from
Figure 11-32 QC Test Selection Menu. Figure 11-54 MAG3
Measurement Message – Fraction #1
screen will appear.
11 - 22 ENHANCED TESTS August 07