CAPINTEC, INC. CRC
®
-25R
DAILY TEST
The Daily Test consists of:
Auto Zero
Background
Chamber Voltage
Data Check
Accuracy Test
Constancy Test
To perform the Daily Test, press TEST from the Measurement Screen. Figure 8-5 Tests
Menu will appear.
TESTS
1. Daily
2. Chamber Volts
3. Accuracy
4. Enhanced
5. Moly Assay
Figure 8-5 Tests Menu
Select DAILY from Figure 8-5 Tests Menu. Figure 8-6 Auto Zero Instruction screen will
appear.
Auto Zero
The first par
t of the Daily Test is the
Auto Zero.
AUTO ZERO
NO SOURCES
Any Key to Continue
Figure 8-6 Auto Zero Instruction
Remove all sources from the vicinity of the Chamber and press any key (except HOME) to
continue. Figure 8-7 Auto Zero Please Wait Screen will appear until a measurement is
available.
October 13 BACKGROUND AND TESTS 8 - 3