CAPINTEC, INC. CRC
®
-25R
Mo99 Assay
The next step is to measure the eluate in the CapMac or Canister.
ELUATE IN
CAPMAC
Any Key to Continue
Figure 10-14 Moly Assay Instruction Screen
Press any key (except HOME) to begin the measurement. Figure 10-15 Moly Assay Wait
Screen will appear until a measurement is available.
Since this is also a measurement of a very low activity, the screens are similar to
background:
PLEASE WAIT
Figure 10-15 Moly Assay Wait Screen
When the measurement is ready, Figure 10-16 Moly Assay Accept Screen will appear.
.47µCi
OK
ENTER to Accept
Figure 10-16 Moly Assay Accept Screen
Press ENTER to accept the measurement and continue to the Tc99m Assay. Figure 10-17
Moly Assay Tc99m Measurement will appear.
March 15 MEASUREMENT PROCEDURES 10 - 13