CAPINTEC, INC. CRC
®
-25R
Zero activity is indicated with a source in the well.
System may not be in lowest range. Try pressing SPC.
Readings appear overly noisy for low activities.
Make sure that the chamber is on a solid surface and is not subject to vibration.
Indication of significant negative activity.
Background level may have changed. Re-do the Background Adjust. Reference
CHAPTER 8: BACKGROUND AND TESTS, SECTION: BACKGROUND.
The time is reset to 00:00.
The internal Lithium Coin Battery may be depleted. Reference the BATTERY
REPLACEMENT procedure on page 15-6.
Printer prints junk or prints with incorrect spacing.
Make sure that the Printer Menu Setup is correct. Reference CHAPTER 5:
SYSTEM INITIALIZATION, SECTION: PRINTING.
Print head hangs up when printing tickets.
On the Okidata Microline 320 printer, the lever on left side of print head must be
set to position 2.
Printer will not respond.
Make sure printer is plugged into a live outlet, turned on, and “selected”.
Make sure that either paper or a ticket is in the paper path.
Run the Diagnostic Test and make sure that the system expects to have a printer.
If not, Make sure that the Printer Menu Setup is correct. Reference CHAPTER 5:
SYSTEM INITIALIZATION, SECTION: PRINTING.
March 15 CLEANING AND MAINTENANCE 15 - 13

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Capintec CRC-25R and is the answer not in the manual?

Capintec CRC-25R Specifications

General IconGeneral
BrandCapintec
ModelCRC-25R
CategoryLaboratory Equipment
LanguageEnglish