CAPINTEC, INC. CRC
®
-25R
Zero activity is indicated with a source in the well.
• System may not be in lowest range. Try pressing SPC.
Readings appear overly noisy for low activities.
• Make sure that the chamber is on a solid surface and is not subject to vibration.
Indication of significant negative activity.
• Background level may have changed. Re-do the Background Adjust. Reference
CHAPTER 8: BACKGROUND AND TESTS, SECTION: BACKGROUND.
The time is reset to 00:00.
• The internal Lithium Coin Battery may be depleted. Reference the BATTERY
REPLACEMENT procedure on page 15-6.
Printer prints junk or prints with incorrect spacing.
• Make sure that the Printer Menu Setup is correct. Reference CHAPTER 5:
SYSTEM INITIALIZATION, SECTION: PRINTING.
Print head hangs up when printing tickets.
• On the Okidata Microline 320 printer, the lever on left side of print head must be
set to position 2.
Printer will not respond.
• Make sure printer is plugged into a live outlet, turned on, and “selected”.
• Make sure that either paper or a ticket is in the paper path.
• Run the Diagnostic Test and make sure that the system expects to have a printer.
If not, Make sure that the Printer Menu Setup is correct. Reference CHAPTER 5:
SYSTEM INITIALIZATION, SECTION: PRINTING.
March 15 CLEANING AND MAINTENANCE 15 - 13