CAPINTEC, INC. CRC
®
-25R
1. Standard
2. Lineator
3. Calicheck
Figure 6-43 Linearity Test Selection Menu
Setting Up Standard Linearity Test
The Standard Test is performed by tracking the decay of a strong Tc99m source.
Select STANDARD from Figure 6-43 Linearity Test Selection Menu. Figure 6-44 Number of
Tests Entry screen will appear.
ENTER NUMBER
OF TESTS
8 – 12
_
Figure 6-44 Number of Tests Entry
This test can be performed with 8 to 12 measurements.
Input the number of measurements that will be performed and press ENTER.
For each measurement (starting with the 2
nd
measurement), the nominal time (in hours) with
respect to the first measurement that the measurement will be performed is input. The
minimum time (in hours) is 1 and the maximum time (in hours) is 255. The example below
(Figure 6-45 Standard Test – Test Time Entry) is for the second measurement.
6 - 24 CHAMBER INITIALIZATION March 15