CAPINTEC, INC. CRC
®
-25R
Input the new Calibration Number and press ENTER. After the calibration number is entered,
the original and new Calibration Numbers will be displayed for verification as shown in Figure
6-39 New Cal# Verification Screen.
Orig Cal #: 080
New Cal #: 082
OK? Y or N
Figure 6-39 New Cal# Verification Screen
If the new Calibration Number is not correct, press NO. Figure 6-38 Enter Calibration
Number screen will re-appear.
If the new Calibration Number is correct, press YES. Figure 6-36 Choose Nuclide for
Calibration Number screen will re-appear.
Restoring Original Calibration Numbers
Choose the nuclide whose Calibration Number is to be restored via the NUCL key, Pre-Set
Nuclide key or USER key. Figure 6-40 Restore Cal#? will appear asking if the original
calibration number should be restored. (Tc99m is shown in the example with 082 as the new
Calibration Number.)
Orig Cal #: 080
New Cal #: 082
Restore Orig Cal # ?
Figure 6-40 Restore Cal#?
To keep the current Calibration Number, press NO. Figure 6-36 Choose Nuclide for
Calibration Number screen will re-appear.
To restore the original Calibration Number, press YES. Figure 6-36 Choose Nuclide for
Calibration Number screen will re-appear.
6 - 22 CHAMBER INITIALIZATION March 15