CAPINTEC, INC CRC
®
-25R
CHAPTER 13
DOSE TABLE
The Dose Table function is used to print a table showing the volume to be withdrawn from a
measured sample to obtain a desired dose at future times for the activity that is being
measured..
Note: A printer must be attached to the system to utilize the Dose Table function.
The Dose Table is accessed by pressing the TIME key twice. It is accessible only for Pre-Set
Nuclide keys (This does not work with the F18, Xe133 keys or any of the USER keys).
Figure
13-1 Dose Entry Screen
will appear.
Enter Dose
value unit
_ mCi
-> to change unit
ENTER to Accept
Figure 13-1 Dose Entry Screen
Input the desired dose and then press the RIGHT ARROW (Ô) key. Use the arrow keys to
scroll through the activity units: mCi, μCi, or Ci for Curie mode (MBq or GBq for Becquerel
mode). If the value needs to be changed, press CE.
Note: The displayed unit (Ci or Bq) is set outside of the Dose Table function. Reference
CHAPTER 5: SYSTEM INITIALIZATION; SECTION: CHOOSING Ci or Bq.
When the value and unit are correct, press ENTER.
Figure 13-2 Volume Entry Screen will
appear.
Enter
Volume In ml
_
Figure 13-2 Volume Entry Screen
July 07 DOSE TABLE 13 - 1