CAPINTEC, INC CRC
®
-25R
Input the volume of the sample that the dose will be drawn from in ml.
Note: The minimum value that can be input is 1.000. The maximum value that can be input
is 100.0.
A report will be printed giving the time and required volume for the desired dose at time
intervals appropriate to the nuclide for up to 30 time intervals. If there is not enough volume
for the dose, the printout will read “REQUIRED VOLUME GREATER THAN EXISTING
VOLUME”. A sample printout is shown in
Figure 13-3 Dose Table Printout.
Figure 13-3 Dose Table Printout
13 - 2 DOSE TABLE July 07