CAPINTEC, INC CRC
®
-25R
Enter
ADDED
Volume In ml
_
Figure 11-10 Added Volume Entry
A minimum of 2 samples must be measured.
Input the first added volume (up to 4 digits) to the vial and press ENTER.
Figure 11-9
Geometry Measurement Screen
will re-appear.
After the 2
nd
and subsequent measurements, Figure 11-11 Another Sample Question
Screen
will appear asking if another sample will be measured. The maximum number of
samples is 10.
Another Sample?
Yes or No
Figure 11-11 Another Sample Question Screen
If the previous sample was not the last sample to be measured, press YES to measure
another sample.
Figure 11-10 Added Volume Entry will re-appear.
Input the added volume (up to 4 digits) for the next sample and press ENTER. The total
volume input must be less than 99ml; if it is not less, the message “VOLUME TOO LARGE –
ANY KEY TO CONTINUE” will appear.
Press NO if the previous sample was the last sample to measured.
Figure 11-12 Geometry
Test Results Screen
will appear.
Measurement Results
After the last sample has been measured,
Figure 11-12 Geometry Test Results Screen will
appear.
August 07 ENHANCED TESTS 11 - 5