CAPINTEC, INC. CRC
®
-25R
For multiplication, the number can only be multiplied by 10 or 100. For division, the number
can only be divided by 2. Refer to Table 6-1 Calibration Number Limits Table on page 6-16.
To input a new Calibration Number, change a current Calibration Number or restore the
original Calibration Number:
Press MENU from the Measurement Screen.
Select SETUP. The Setup Menu appears.
Select OTHER. Input the password (last 3 digits of Readout serial number) and press
ENTER. Figure 6-1 Other Menu appears.
From the Other Menu, select NUCLIDES. Figure 6-24 Nuclides Menu will appear.
From the Nuclide Menu, select NEW CAL NUM. Figure 6-32 Choose Nuclide for Calibration
Number screen will appear.
CHOOSE NUCLIDE
Press NUCL or
Pre-set nuclide key
or User key
Figure 6-32 Choose Nuclide for Calibration Number
The nuclide can be chosen via the NUCL key, Pre-Set Nuclide key or USER key.
Entering New Calibration Numbers
If a the selected nuclide does not have a Calibration Number, Figure 6-33 Change Original
Cal#? Screen will appear.
Orig Cal #: NONE
Change Cal #?
Figure 6-33 Change Original Cal#?
March 15 CHAMBER INITIALIZATION 6 - 19