CAPINTEC, INC CRC
®
-25R
Input the time as hhmm in 24 hour time format (example: 1:25 PM would be 1325, 9:15 AM
would be 0915) and press ENTER. The time is checked for validity. If the input time is not
valid, a beep will sound and “TIME ERROR” will be displayed. The time must be re-entered.
The input activity, the date and time of use, and the volume required for that activity will be
displayed. The prompt “OK? Y or N” will appear. Press NO to return to the
Figure 12-22
Required Activity Entry Screen
.
If the displayed information is correct, withdraw the displayed volume and press YES. Figure
12-25 Measure Activity Screen
will appear.
MEASURE
ACTIVITY
Any Key to Continue
Figure 12-25 Measure Activity Screen
Place the sample in the Chamber and press any key to continue.
Figure 12-26 Withdraw
Activity Measurement Screen
will appear.
Tc99m
6.27mCi
ENTER to Accept
Figure 12-26 Withdraw Activity Measurement Screen
When the activity has stabilized, press ENTER to accept the measurement. The Inventory
item will update with the new volume and activity and Mo to Tc ratio for Tc99m (unless 0 was
entered for the ratio).
Figure 12-2 Inventory Menu will re-appear.
PRINTING THE INVENTORY
If a printer is attached to the system, the Inventory can be printed.
To print the Inventory, select PRINT from
Figure 12-2 Inventory Menu.
12 - 12 INVENTORY August 07