CAPINTEC, INC CRC
®
-25R
The first measurement made is considered the Base Measurement. The base measurement
is the measurement that all other measurements will be compared to. The variance is
calculated using the base measurement.
GEOMETRY TEST
Using: Syringe
Volume Assay %Var
10.0ml 10.9mCi BASE
11.0ml 11.0mCi 0.9
12.0ml 10.8mCi -1.8
An
y
Ke
y
to Continue
Figure 11-12 Geometry Test Results Screen
Press any key (except HOME) to continue. If a printer is attached to the system, a report will
be printed and the prompt to print again will appear.
To print the test results again, press YES. To return to
Figure 11-1 Tests Menu, press NO.
The report will show the variation of each measurement with respect to the selected base.
The variation for the base will be shown as BASE.
PERFORMING LINEARITY TEST
Note: In order to perform a Linearity Test, a Linearity Test Method must be defined from the
Linearity Setup in the Other Menu. (reference CHAPTER 6: CHAMBER
INITIALIZATION, SECTION: LINEARITY TEST DEFINITION), If a Linearity Method is
not defined,
Figure 11-13 Linearity Not Defined Error Screen will appear.
Must Define
Linearity Test
Any Key to Continue
Figure 11-13 Linearity Not Defined Error Screen
To perform a Linearity Test, press TEST from the Measurement Screen.
Figure 11-1 Tests
Menu
will appear.
Select ENHANCED.
Figure 11-2 Enhanced Tests Menu will appear.
Select LINEARITY. The defined linearity test will begin.
11 - 6 ENHANCED TESTS August 07