Page #136 background imageLoading...
Page #136 background image
CAPINTEC, INC CRC
Ā®
-25R
Press any key (except HOME) to continue to the next Results screen.
CALICHECK TEST
MEAN: n.nnnmCi
# RESULT % VAR
5 x.xxxmCi x.xx
6 x.xxxmCi x.xx
7 x.xxxmCi x.xx
8 x.xxxmCi x.xx
An
y
Ke
y
to Continue
Figure 11-31 Calicheck – 2
nd
Test Results Screen
If more than 8 tests have been selected in the test setup, a 3
rd
results screen will appear
showing the values for the 9
th
through last test.
Press any key (except HOME) to continue. If a printer is attached to the system, a Calicheck
Linearity Test report will print. The report will show the result and the variation from mean for
each tube.
The prompt to print again will appear. To print the test results again, press YES. To return to
Figure 11-1 Tests Menu, press NO.
QUALITY CONTROL (QC)
Radiochromatography Quality Control tests can be performed using four methods.
Press TEST from the Measurement Screen.
Figure 11-1 Tests Menu will appear.
Select ENHANCED.
Figure 11-2 Enhanced Tests Menu will appear.
Select QC.
Figure 11-32 QC Test Selection Menu will appear.
QC TESTS
1. 1 Strip
2. 2 Strips
3. HMPAO
4. MAG3
Figure 11-32 QC Test Selection Menu
11 - 14 ENHANCED TESTS August 07

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Capintec CRC-25R and is the answer not in the manual?

Capintec CRC-25R Specifications

General IconGeneral
BrandCapintec
ModelCRC-25R
CategoryLaboratory Equipment
LanguageEnglish