Place the requested tube(s) in the Chamber and press any key (except HOME) to perform
the specified measurement.
Figure 11-28 Calicheck Measurement Screen will appear.
Tc99m
6.27mCi
ENTER to Accept
Figure 11-28 Calicheck Measurement Screen
Press ENTER to accept the measurement.
Figure 11-29 Calicheck Measurement
Verification
screen will appear:
Tube # 5
Black + Green
Result n.nnnmCi
OK? Y or N
Figure 11-29 Calicheck Measurement Verification
To repeat the measurement or if the measurement is not acceptable, press NO.
Figure 11-27
Calicheck Measurement Prompt
will re-appear.
If the measurement is acceptable, press YES. The next Calicheck Measurement Prompt
screen will appear until all tubes are measured.
After the last measurement, the results screens will appear.
CALICHECK TEST
MEAN: n.nnnmCi
# RESULT % VAR
1 x.xxxmCi x.xx
2 x.xxxmCi x.xx
3 x.xxxmCi x.xx
4 x.xxxmCi x.xx
An
Ke
o Continue
Figure 11-30 Calicheck ā 1
st
Test Results Screen
August 07 ENHANCED TESTS 11 - 13