CAPINTEC, INC. CRC
®
-25R
QUARTERLY TESTS
Diagnostic Test
The Diagnostics Test should be performed as a part of the Quarterly Tests. When
Diagnostics is selected, the instrument's memories and programs are checked and the
results are displayed. If a printer is attached to the system, the results can be printed. If any
of the tests fail, contact Capintec’s only Authorized Service Center at 1-800-227-6832.
(Reference CHAPTER 9: DIAGNOSTICS).
YEARLY TESTS
Linearity Test
The linearity of the CRC
®
-25R should be checked over the entire range of activities which are
reasonably anticipated to be used. The initial linearity should be performed using decay
method of measuring a short lived isotope over time. Thereafter, use of calibrated sleeves is
acceptable.
The three most common methods are described in CHAPTER 6: CHAMBER
INITIALIZATION, SECTION: LINEARITY TEST DEFINITION and CHAPTER 11:
ENHANCED TESTS, SECTION: PERFORMING LINEARITY TEST.
7 - 4 ACCEPTANCE & QUALITY ASSURANCE TESTS March 15