CAPINTEC, INC. CRC
®
-25R
SELECT
1. CAPMAC for
Mallinckrodt Gen
2. CAPMAC for
Bristol Myers Gen
3. Capintec Canister
Figure 6-18 Moly Assay Method Selection
Press MENU from the Measurement Screen.
Select SETUP. The Setup Menu appears.
Select OTHER. Input the password (last 3 digits of Readout serial number) and press
ENTER. Figure 6-1 Other Menu appears.
From the Other Menu, select MOLY SETUP. Figure 6-19 Moly Setup Menu will appear.
MO ASSAY
1. Moly Method
2. Mo/Tc Limit
Figure 6-19 Moly Setup Menu
Selecting Moly Assay Method
Select MOLY METHOD from the Moly Setup Menu. Figure 6-20 Current Moly Assay
Methods Screen will appear.
The current selections are displayed (default is all are selected):
March 15 CHAMBER INITIALIZATION 6 - 11