CAPINTEC, INC CRC
®
-25R
CONVERSION BETWEEN Ci and Bq (Ci Bq)
An activity is input in Curies and the value is displayed in Becquerels or an activity is input in
Becquerels and the value is displayed in Curies.
From
Figure 14-2 Calculations Menu, select Ci
Bq CONV. Figure 14-3 Ci Bq Activity
Entry Screen
will appear.
ENTER ACTIVITY
value unit
_ µCi
-> to change unit
ENTER to Accept
Figure 14-3 Ci Bq Activity Entry Screen
Input the value of the activity to be converted and then press the RIGHT ARROW (Ô) key.
Use the arrow keys to scroll the activity units: µCi, GBq, MBq, Ci, mCi. If the value needs to
be changed, press CE.
When the activity value and units are correct, press ENTER.
Figure 14-4 Ci Bq Result
Screen
will appear (in the example 100 mCi was entered):
100.0mCi
` 3.70GBq
ENTER to Continue
MENU to End
Figure 14-4 Ci Bq Result Screen
To perform another conversion, press ENTER.
Figure 14-3 Ci Bq Activity Entry Screen
will re-appear.
To exit the conversion utility, press MENU.
Figure 14-2 Calculations Menu will re-appear.
14 - 2 CALCULATION UTILITIES July 07