CAPINTEC, INC CRC
®
-25R
Note: The minimum value that can be input is 1.000. The maximum value that can be input
is 100.0.
MEASURE
ACTIVITY ?
Yes or No
Figure 12-8 Measure Activity Question Screen
The activity may be measured or it may be input if activity and time of a previous
measurement are known.
To input the activity and time, press NO.
Figure 12-9 Inventory Activity Entry Screen will
appear.
To measure the activity, press YES.
Figure 12-12 Inventory Measurement Screen will
appear.
Input Activity
If the activity is to be input,
Figure 12-9 Inventory Activity Entry Screen will appear:
ENTER ACTIVITY
value unit
_ mCi
-> to change unit
ENTER to Accept
Figure 12-9 Inventory Activity Entry Screen
Input the value of the activity and then press the RIGHT ARROW (Ô). Use the arrow keys to
scroll the activity units: mCi, µCi, Ci if using Curie mode, (or MBq, GBq if using Becquerel
mode). If the value needs to be changed, press CE.
When the activity value and units are correct, press ENTER.
Figure 12-10 Inventory Date
Entry Screen
will appear.