CHAPTER 15
CLEANING AND MAINTENANCE
GENERAL
This chapter provides the information necessary for the user to perform the basic
maintenance of instrument cleaning, fuse replacement, battery replacement and general
preventative maintenance. There are no internal adjustments or calibration settings that may
be done by the user within the conditions of the warranty.
CAUTION: REFER ALL SERVICING TO A QUALIFIED SERVICE
REPRESENTATIVE!
It is recommended that periodic (every five years) re-calibration of the CRC
®
-25R be
performed only by Capintec’s Authorized Service Center to guarantee the instrument’s high
reliability is maintained. Contact Capintec’s only Authorized Service Center in Pittsburgh for
servicing or re-calibration at 1-800-227-6832.
CLEANING and DISINFECTING
CAUTION:
DISCONNECT THE POWER BEFORE CLEANING.
TO AVOID ELECTRICAL SHOCK OR DAMAGING OF THE CRC
®
-25R,
NEVER ALLOW WATER OR LIQUIDS TO PENETRATE THE CHAMBER OR
THE READOUT ENCLOSURE.
DO NOT USE AEROSOL DISPENSERS TO SPRAY THE EQUIPMENT WITH
CLEANING SOLUTIONS OR LIQUIDS.
TO AVOID DAMAGING THE CASE OR DISPLAY SCREEN, DO NOT USE
AROMATIC HYDROCARBONS, CHLORINATED SOLVENTS OR
METHANOL-BASED CLEANING SOLUTIONS.
PRIOR TO CLEANING OR DISINFECTING THE LINER AND/OR DIPPER,
THEY MUST BE REMOVED FROM THE CHAMBER. CAUTION: NEVER USE
THE CALIBRATOR WITHOUT THE CHAMBER LINER IN PLACE. LINERS
ARE INEXPENSIVE AND EASY TO REPLACE. A CONTAMINATED
CHAMBER IS A VERY COSTLY MISTAKE.
Cleaning Instructions
Readout Unit and Chamber
Wipe the surfaces clean using a damp, non-abrasive cloth or sponge and a mild
detergent and water; do not use solvents or aerosol cleaners. After cleaning, wipe all
surfaces dry with a soft, non-abrasive cloth. To avoid scratches, do not use abrasive
pads.
March 15 CLEANING AND MAINTENANCE 15 - 1