CAPINTEC, INC. CRC
®
-25R
LINEARITY TEST
TIME FOR
MEASUREMENT 2
IN HOURS
Figure 6-45 Standard Test – Test Time Entry
Input the time (in hours) and press ENTER. The above screen will be repeated for each
measurement.
After all the times are input, the Standard Test Verification screen will appear. The example
shown (Figure 6-46 Standard Test Verification) is for 12 tests at 1 hour increments:
1: 0 2: 1 3: 2
4: 3 5: 4 6: 5
7: 6 8: 7 9: 8
10: 9 11: 10 12: 11
OK? Y or N
Figure 6-46 Standard Test Verification
If the displayed times are incorrect, press NO. Figure 6-44 Number of Tests Entry will re-
appear.
If the displayed times are correct, press YES. Figure 6-1 Other Menu will reappear.
Setting Up Lineator Test
This test requires 8 measurements. A Lineator Test Calibration procedure must be
performed before any Lineator tests.
Select LINEATOR from Figure 6-43 Linearity Test Selection Menu. Figure 6-47 Lineator
Calibration Prompt screen will appear. The example below is given for the 5
th
measurement.
March 15 CHAMBER INITIALIZATION 6 - 25