CAPINTEC, INC. CRC
®
-25R
Measure Tube # 5
1 + 4
Any Key to Continue
Figure 6-47 Lineator Calibration Prompt
Place the Linearity Test Source in the calibrator and press any key to continue the
measurement. Figure 6-48 Lineator Measurement Screen will appear.
6.27mCi
ENTER to Accept
Figure 6-48 Lineator Measurement Screen
Press ENTER to accept the measurement. Figure 6-49 Lineator Calibration Measurement
Verification screen will appear.
1 + 4
Activity: x.xxmCi
Initial Factor
n.nn
OK? Y or N
Figure 6-49 Lineator Calibration Measurement Verification
To repeat the measurement, press NO. Figure 6-47 Lineator Calibration Prompt will re-
appear.
To accept the measurement, press YES. The next Lineator Calibration Prompt screen will
appear until all 8 tubes are measured.
If a printer is attached to the system, a Lineator Test Calibration report will print. The report
will show the measured activity and initial factor for each tube set.
6 - 26 CHAMBER INITIALIZATION March 15