EasyManuals Logo

Capintec CRC-25R User Manual

Capintec CRC-25R
228 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #130 background imageLoading...
Page #130 background image
CAPINTEC, INC CRC
®
-25R
Measured x.xxmCi
OK? Y or N
Figure 11-16 Standard Linearity Test First Measurement Verification
To repeat the measurement or if the measurement is not acceptable, press NO.
Figure 11-14 Standard Linearity Test First Measurement Prompt will re-appear.
If the measurement is acceptable, press YES. If a printer is attached to the system, a
Standard Linearity Test First Measurement report will print and the prompt to print
again will appear.
To print the test results again, press YES. To return to
Figure 11-1 Tests Menu,
press NO.
The report will show the predicted and actual activity and the variance.
Second Through Last Measurements
Perform the measurements at time intervals close to that entered when the test was
defined. The actual elapsed time will be used in the calculations.
From
Figure 11-1 Tests Menu, select ENHANCED. Figure 11-2 Enhanced Tests
Menu
will appear.
Select LINEARITY. Figure 11-17 Standard Linearity Test – Previous Test Results
Screen
will appear.
In the example, 2 measurements have been made.
STANDARD LINEARITY
# MEASURED %VAR
1 x.xxmCi x.xx
2 x.xxmCi x.xx
New Test? Y or N
Figure 11-17 Standard Linearity Test – Previous Test Results Screen
11 - 8 ENHANCED TESTS August 07