EasyManua.ls Logo

Capintec CRC-25R - Dimensions; Console; Chamber; Cables

Capintec CRC-25R
228 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CAPINTEC, INC. CRC
®
-25R
Dimensions
Console
Height ........................ 13.7cm ..... (5.38in)
Width ......................... 26.0cm ..... (10.25in)
Depth ........................ 26.7cm ..... (10.5in)
Weight ....................... 1.8kg ........ (3.9lb)
Chamber
Height ........................ 43.8cm ..... (17.25in)
Diameter .................... 17.2cm ..... (6.76in)
Weight ....................... 13.6kg ...... (30lb)
Well Diameter ............ 6.1cm ....... (2.4in)
Well Depth ................. 25.4cm ..... (10.0in)
Cable Length
1
............ 3.7m ........ (12ft)
Cables
Power ........................ 1.8m ........ (6ft)
Printer
2
...................... 1.8m ........ (6ft)
Performance
Measurement Range:
Maximum Activity (Co60) .................. 37.4 GBq (1.00 Ci)
Maximum Activity (Co57) .................. 206. GBq (5.57 Ci)
Resolution .................................................... 0.001 MBq (.01 µCi)
Electrometer Accuracy
3
................................ better than ±2%
System Precision ......................................... better than ± 0.1% of FSD
System Linearity .......................................... within ±2%
Response Time
Below 20µCi ..................................... within 25 seconds
Above 20µCi ..................................... within 4 seconds
1
Longer cables are available. Consult factory.
2
Optional
3
Overall accuracy is determined by the calibration for the specific nuclide and the sample configuration and the accuracies of
the standard sources used for calibration of the electrometer.
2 - 6 FUNCTIONAL & TECHNICAL DESCRIPTION March 15