Dimensions
Console
Height ........................ 13.7cm ..... (5.38in)
Width ......................... 26.0cm ..... (10.25in)
Depth ........................ 26.7cm ..... (10.5in)
Weight ....................... 1.8kg ........ (3.9lb)
Chamber
Height ........................ 43.8cm ..... (17.25in)
Diameter .................... 17.2cm ..... (6.76in)
Weight ....................... 13.6kg ...... (30lb)
Well Diameter ............ 6.1cm ....... (2.4in)
Well Depth ................. 25.4cm ..... (10.0in)
Cable Length
1
............ 3.7m ........ (12ft)
Cables
Power ........................ 1.8m ........ (6ft)
Printer
2
...................... 1.8m ........ (6ft)
Performance
Measurement Range:
Maximum Activity (Co60) .................. 37.4 GBq (1.00 Ci)
Maximum Activity (Co57) .................. 206. GBq (5.57 Ci)
Resolution .................................................... 0.001 MBq (.01 µCi)
Electrometer Accuracy
3
................................ better than ±2%
System Precision ......................................... better than ± 0.1% of FSD
System Linearity .......................................... within ±2%
Response Time
Below 20µCi ..................................... within 25 seconds
Above 20µCi ..................................... within 4 seconds
1
Longer cables are available. Consult factory.
2
Optional
3
Overall accuracy is determined by the calibration for the specific nuclide and the sample configuration and the accuracies of
the standard sources used for calibration of the electrometer.