Source Verification
If the chosen source had data previously entered, the data will be displayed. The example
below is for Co57, which will be used in the Daily Test:
Co57
S/N: 1458
JAN 07 2006
100.0 mCi DAILY
OK? Y or N
Figure 6-7 Test Source Verification
If the data is correct, press YES. Figure 6-5 Test Source Menu will re-appear.
To change the data or to delete the source, press NO. Figure 6-8 Source Existence screen
will appear.
Do You Have
Co57 ?
Yes or No
Figure 6-8 Source Existence
If the selected Test Source is no longer being used, press NO and it will be deleted. Figure 6-
5 Test Source Menu will re-appear.
To input data for the selected Test Source, press YES. Figure 6-9 Entering Serial Number
screen will appear.