EasyManua.ls Logo

Capintec CRC-25R - Page 78

Capintec CRC-25R
228 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CAPINTEC, INC. CRC
®
-25R
To leave the Calibration Number set to NONE, press NO. Figure 6-32 Choose Nuclide for
Calibration Number screen will re-appear.
To input the new Calibration Number, press YES. Figure 6-34 Enter Calibration Number
screen will appear.
ENTER
CALIBRATION #
_
Figure 6-34 Enter Calibration Number
Input the new Calibration Number and press ENTER. After the Calibration Number is
entered, Figure 6-35 New Cal# Verification Screen will appear.
Sr89
Orig Cal #: NONE
New Cal #: 123
OK? Y or N
Figure 6-35 New Cal# Verification Screen
If the new Calibration Number is not correct, press NO. Figure 6-34 Enter Calibration
Number screen will re-appear.
If the new Calibration Number is correct, press YES. The new Calibration Number will be
stored and Figure 6-32 Choose Nuclide for Calibration Number screen will re-appear.
6 - 20 CHAMBER INITIALIZATION March 15

Table of Contents