CAPINTEC, INC. CRC
®
-25R
BACKGROUND
PLEASE WAIT
Figure 8-2 Background Please Wait Screen
When a measurement is available, Figure 8-3 Background Results Screen will appear.
BACKGROUND
0.56µCi
OK
ENTER to Accept
Figure 8-3 Background Results Screen
If the background is high but still acceptable (> 10µCi [0.4MBq] and < 530µCi [20MBq]) the
message “HIGH” will appear instead of “OK”. Although the value is acceptable, the reason
for the high value should be investigated. If any sources are found nearby, repeat the
measurement.
If the background is above the acceptable range (> 530µCi [20MBq]), the value will not be
shown but Figure 8-11 Background Too High Screen will appear.
BACKGROUND
TOO HIGH
SEE MANUAL
Any key to Continue
Figure 8-4 Background Too High Screen
This “TOO HIGH” background cannot be accepted by the CRC
®
-25R. If the cause of the high
reading (nearby source, contaminated well etc.) cannot be found, contact Capintec’s only
Authorized Service Center at 1-800-227-6832.
Press ENTER to accept the result. The Measurement Screen will re-appear.
8 - 2 BACKGROUND AND TESTS October 13