CAPINTEC, INC CRC
®
-25R
Place Fraction #1
in Chamber
Any Key to Continue
Figure 11-54 MAG3 Measurement Message – Fraction #1
Place the Fraction #1 in the Chamber and press any key (except HOME) to perform the
measurement.
Figure 11-55 QC Test Measurement Screen will appear.
Tc99m
6.27mCi
ENTER to Accept
Figure 11-55 QC Test Measurement Screen
Press ENTER to accept the measurement.
Figure 11-56 MAG3 Measurement Message –
Fraction #2
screen will appear.
Place Fraction #2
in Chamber
Any Key to Continue
Figure 11-56 MAG3 Measurement Message – Fraction #2
Place the Fraction #2 in the Chamber and press any key (except HOME) to perform the
measurement.
Figure 11-55 QC Test Measurement Screen will re-appear.
Press ENTER to accept the measurement.
Figure 11-57 MAG3 Measurement Message –
Cartridge
screen will appear:
August 07 ENHANCED TESTS 11 - 23