CAPINTEC, INC CRC
®
-25R
Place Cartridge
in Chamber
Any Key to Continue
Figure 11-57 MAG3 Measurement Message – Cartridge
Place the Cartridge in the Chamber and press any key (except HOME) to perform the
measurement.
Figure 11-55 QC Test Measurement Screen will re-appear.
Press ENTER to accept the measurement.
Figure 11-58 MAG3 Results Screen will appear.
QC TEST MAG3
Non-elutable Tc99m
xx.x%
Hydrophilic Impure
xx.x%
Tc99m Mertiatide
xx.x%
An
Ke
to Continue
Figure 11-58 MAG3 Results Screen
Press any key (except HOME) to continue. If a printer is attached to the system, a QC Test
MAG3 report will print. The report will show the measured values and results.
The prompt to print again will appear. To print the test results again, press YES. Press NO to
continue.
Figure 11-59 Perform Another Test Screen will appear.
Do Another
Test? Y or N
Figure 11-59 Perform Another Test Screen
To perform another test, press YES.
Figure 11-54 MAG3 Measurement Message – Fraction
#1
will re-appear.
If another test is not desired, press NO to exit.
Figure 11-1 Tests Menu will appear.
11 - 24 ENHANCED TESTS August 07