CAPINTEC, INC CRC
®
-25R
Place TOP of
Strip A in Chamber
Any Key to Continue
Figure 11-38 2 Strips Measurement Message – Top A
Place the Top of Strip A in the Chamber and press any key (except HOME) to perform the
measurement.
Figure 11-39 QC Test Measurement Screen will appear.
Tc99m
6.27mCi
ENTER to Accept
Figure 11-39 QC Test Measurement Screen
Press ENTER to accept the measurement.
Figure 11-40 2 Strips Measurement Message –
Bottom A
screen will appear.
Place BOTTOM of
Strip A in Chamber
Any Key to Continue
Figure 11-40 2 Strips Measurement Message – Bottom A
Place the Bottom of Strip A in the Chamber and press any key (except HOME) to perform the
measurement.
Figure 11-39 QC Test Measurement Screen will appear.
Press ENTER to accept the measurement.
Figure 11-41 2 Strips Measurement Message –
Top B
screen will appear.
August 07 ENHANCED TESTS 11 - 17