CAPINTEC, INC CRC
®
-25R
Place TOP of
Strip B in Chamber
Any Key to Continue
Figure 11-41 2 Strips Measurement Message – Top B
Place the Top of Strip B in the Chamber and press any key (except HOME) to perform the
measurement.
Figure 11-39 QC Test Measurement Screen will re-appear.
Press ENTER to accept the measurement.
Figure 11-42 2 Strips Measurement Message –
Bottom B
screen will appear:
Place BOTTOM of
Strip B in Chamber
Any Key to Continue
Figure 11-42 2 Strips Measurement Message – Bottom B
Place the Bottom of Strip B in the Chamber and press any key (except HOME) to perform the
measurement.
Figure 11-39 QC Test Measurement Screen will re-appear.
Press ENTER to accept the measurement. Figure 11-43 2 Strips Results Screen will appear.
QC TEST Two Strips
Free Tc99m
xx.x%
Reduced/Hydrolized
xx.x%
Purity
xx.x%
An
y
Ke
y
to Continue
Figure 11-43 2 Strips Results Screen
Press any key (except HOME) to continue. If a printer is attached to the system, a QC Test
2 Strip report will print. The report will show the measured values and results.
11 - 18 ENHANCED TESTS August 07

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Capintec CRC-25R and is the answer not in the manual?

Capintec CRC-25R Specifications

General IconGeneral
TypeDose Calibrator
ManufacturerCapintec
ModelCRC-25R
DisplayLCD
Chamber Well Diameter2.5" (6.35 cm)
Accuracy± 5%
Response Time1 second
CalibrationAutomatic or manual
Nuclide LibraryOver 100 nuclides
ConnectivityRS232, USB
Power100-240 VAC, 50/60 Hz