CAPINTEC, INC CRC
®
-25R
Tc99m
6.27mCi
ENTER to Accept
Figure 11-46 QC Test Measurement Screen
Press ENTER to accept the measurement.
Figure 11-47 HMPAO Measurement Message –
Bottom A
screen will appear.
Place BOTTOM of
Strip A in Chamber
Any Key to Continue
Figure 11-47 HMPAO Measurement Message – Bottom A
Place the Bottom of Strip A in the Chamber and press any key (except HOME) to perform the
measurement.
Figure 11-46 QC Test Measurement Screen will re-appear.
Press ENTER to accept the measurement.
Figure 11-48 HMPAO Measurement Message –
Top B
screen will appear.
Place TOP of
Strip B in Chamber
Any Key to Continue
Figure 11-48 HMPAO Measurement Message – Top B
Place the Top of Strip B in the Chamber and press any key (except HOME) to perform the
measurement.
Figure 11-46 QC Test Measurement Screen will re-appear.
Press ENTER to accept the measurement. Figure 11-49 HMPAO Measurement Message –
Bottom B
screen will appear:
11 - 20 ENHANCED TESTS August 07