EasyManua.ls Logo

Capintec CRC-25R - Page 207

Capintec CRC-25R
228 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CAPINTEC, INC CRC
®
-25R
Uncertainty Due to
Radioisotopes
Calibration
Setting
Number
Syringe
Corr. %
Published
Data %
Half-Life
(NCRP-58)
Ref. Comments
204
Tl Thallium
420 × 100
2 2 3.78 Y 2 NDT70
207
Bi Bismuth 846 1.7 32.2 Y 9 NBS73 Ref. for 1064, 569.7,
76.7, & 1772 keV.
208
Tl Thallium
571 ÷ 2
3.053 M 4 NM75
212
Pb Lead 101 10.64 H 1 NM75 Decays to
212
Bi. Eqb.
after 1 hr. See App. IV.
212
Bi Bismuth
489 × 10
60.55 M 6 NM75
212
Pb Lead
(Eqb.
212
Bi)
158 Reading gives
212
Pb
Act. in eqb. sample.
Eqb. after 8 hours.
212
Bi Bismuth
(Eqb.
212
Pb)
135 Reading gives
212
Bi
Act. in eqb. sample.
212
Pb
212
Bi
Lead
Eqb.
Bismuth
030 or
146 × 2
Reading gives sum of
212Pb & 212Bi activity
in equilibrium sample.
224
Ra Radium
646 × 100
3.66 D 4 NM75
226
Ra Radium
+ chain of
daughters
778 0.5 1600 Y 7 NBS73 Reading in grams.
Commonly referred to
as "Radium "Source.
1.025 g/Ci of Ra-226.
239
Np Neptunium 147 6 2.355 D 4 ORNL76
241
Am Americium 055 4 1 432.2 Y 5 LMR69 Ref. for 59.5 & 14 keV.
June 09 APPENDIX II A2 - 15

Table of Contents