EasyManua.ls Logo

Capintec CRC-25R - MULTIPLICATION FACTORS FOR CA47 ACTIVITY MEASUREMENTS; MULTIPLICATION FACTORS FOR NI56 ACTIVITY MEASUREMENTS

Capintec CRC-25R
228 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CAPINTEC, INC CRC
®
-25R
Multiplication factors for Ca47 Activity Measurements:
Use a Calibration Number of 373 and a multiplication factor as indicated in the following table
to obtain Ca47 activity at time “t” days after an initially pure source of Ca47 is obtained.
Ca47
t
[days]
Multiplication
Factor
t
[days]
Multiplication
Factor
0 1.000 12 0.672
1 0.950 14 0.648
2 0.906 16 0.627
3 0.869 18 0.609
4 0.836 20 0.594
5 0.806 25 0.566
6 0.781 30 0.545
7 0.757 40 0.519
8 0.737 50 0.496
9 0.718 100 0.485
10 0.701 150 0.484
Multiplication factors for Ni56 Activity Measurements:
Use a Calibration Number of 844 and a multiplication factor as indicated in the following table
to obtain Ni56 activity at time “t” days after an initially pure source of Ni56 is obtained.
Ni56
t
[days]
Multiplication
Factor
t
[days]
Multiplication
Factor
0 1.000 9 0.824
1 0.985 10 0.799
2 0.969 12 0.745
3 0.952 15 0.659
4 0.934 20 0.508
5 0.915 25 0.366
6 0.894 30 0.249
7 0.872 40 0.101
8 0.849 60 0.013
A3 - 2 APPENDIX III July 07

Table of Contents