CAPINTEC, INC. CRC
®
-25R
or Ra226. There can be a Test Source for each of these nuclides. One or more of the
Test Sources must be designated as the source(s) to be used in the Daily Test.
Reference CHAPTER 6: CHAMBER INITIALIZATION, SECTION: TEST SOURCES.
Enter Constancy Test Source: The source for the Constancy Test is chosen from the
sources designated as Daily Sources. Reference CHAPTER 6: CHAMBER
INITIALIZATION, SECTION: CONSTANCY TEST SOURCE.
Verify correct printer setting: The CRC
®
-25R may be used with one of the following
printers. Reference CHAPTER 5: SYSTEM INITIALIZATION, SECTION: PRINTING.
o Okidata Microline 320
o Epson LX-300+II
o Epson TM-U295 SLIP Printer
o Epson TM-U220D Roll Printer
o USB HP Inkjet
o USB Epson Inkjet
The following items should be performed before use if required:
Assign USER Keys: The USER Keys (U1, U2, U3, U4, U5) are a quick way to select
a nuclide. A nuclide may be assigned to each key. Reference CHAPTER 6:
CHAMBER INITIALIZATION, SECTION: USER KEY ASSIGNMENT.
Adding a Nuclide: The CRC
®
-25R contains nuclide data (name, half-life, calibration
number) for over 80 nuclides. The user may add up to 10 nuclides: the name, half-life
and calibration number will be added for each one. Reference CHAPTER 6:
CHAMBER INITIALIZATION, SECTION: ADDING A NUCLIDE.
Changing Calibration Numbers: The user is allowed to change the calibration number
of up to 10 built-in nuclides. Reference CHAPTER 6: CHAMBER INITIALIZATION,
SECTION: CALIBRATION NUMBERS.
Selecting Moly Assay Method: Moly Assay can be performed using the CAPMAC or
CANISTER. Reference CHAPTER 6: CHAMBER INITIALIZATION, SECTION:
SELECTING MOLY ASSAY METHOD.
Setting Mo/Tc Limit: The Mo/Tc limit is set at the factory to 0.15µCi/mCi. The user can
change this value. Reference CHAPTER 6: CHAMBER INITIALIZATION, SECTION:
MOLY ASSAY LIMIT.
March 15 SYSTEM SETUP 4 - 7