Horizon Imager User’s Manual 6-23
Printing via PostScript
Priority Low, Medium, High Priority of job sent to imager. Higher priority jobs are
printed first.
Image Settings (Print dialog box)
Gamma 0.1 to 6.0, in increments
of 0.5
Affects image brightness.
Contrast -100 to 100, in increments
of 5
Affects image contrast.
Polarity Positive, Negative Can be used to reverse image polarity.
Background White/Clear, Light,
Medium, Dark, Black
Sets color of any part of sheet or cell not covered by an
image.
Scaling Method Box, Bilinear, Mitchell,
Bicubic, Lanczos,
Sharp1, Sharp2, Sharp3
Specifies the filter to use if the image is being scaled
(scaling can be specified in the Sheet Layout parameter in
Set 1).
Saturation -100 to 100, in increments
of 5
A property of color that specifies how far a color is from a
gray tone of the same density.
Negative values reduce saturation; positive values
increase saturation.
-100 = no color; 0 = no adjustment; 100 = full color.
MCM 0 to 17, plus other custom
values
Medical Color Matching. For values 1 through 17,
algorithms are applied to the image data before printing so
that printed images match those displayed on a monitor.
For a list of which monitors correspond to the available
MCM values, refer to Appendix D.
Table 6-2. Horizon Imager Macintosh PostScript Parameters (cont.)