Horizon Imager User’s Manual xxix
Preface
WARNING Use of accessories, transducers and cables other than those
specified or provided by the manufacturer of this equipment could result in
increased electromagnetic emissions or decreased electromagnetic immunity
of this equipment and result in improper operation.
WARNING Portable RF communications equipment (including peripherals
such as antenna cables and external antennas) should be used no closer than
30 cm (12 inches) to any part of the Horizon, its cables, or accessories.
Otherwise, degradation of the performance of this equipment could result.
Electromagnetic Emissions Standards and Test Levels:
Test / Standard Compliance Level
RF Emissions
CISPR 11
Group 1, Class B
RF Emissions
FCC Part 15
Class B
Conducted Emissions
CISPR 11
Group 1, Class B
Harmonic Distortion
IEC 61000-3-2
Class B
Voltage Fluctuations and Flicker
IEC 61000-3-3
Complies
Electromagnetic Immunity Standards and Test Levels:
Test / Standard Compliance Level
Electrostatic Discharge
IEC 61000-4-2
±8kV contact
±2kV, ±4kV, ±8kV, ±15kV air