Chapter 3 - Preoperational Setup -
AS-410M Auto Slide Preparation System
77
3-16-2 Setting change of the cooler temperature
Follow the procedure below to change the temperature setting.
Push and hold SET button for five seconds to go to the program mode.
The 7 segment display switches to .
Push SET button again.
The display shows the current target setting temperature. Push button (raise) or push
button (lower) to display .
Push the button (raise) or the button (lower) again to display .
Then push SET button.
Change to the target temperature; to raise the temperature, to lower the temperature,
and change to 20.
Then, again push SET button to display ; the setting temperature is completed.
After the completion of the temperature setting, push SET button and hold 5 seconds to display
The screen indication will switch to the current cooling temperature.