Chapter 5 Errors and Troubleshooting
AS-410M Auto Slide Preparation System
209
5-3-8 Extension Packet errors
In case errors continue to occur after the AS-410M initialization and reboot of the LK-400
application, the file data may be broken or it may be a system failure. Retry the section making with
a sample that has a performance record; and check the data contents.
Error
code
On-screen message Possible cause Solution
9000
<Error Message>
Unit:
Sample feeding unit
Part:
Sample tray
Symptom: No block during manual
input
No block in the specified
position at the start of
sectioning with manual input
mode.
・Check whether there is
a sample block in the
position specified by LK
PC.
9001
<Error Message>
Unit:
Sample feeding unit
Part:
Sample tray
Symptom:
Manual input blocks
present
There is a block at the
unspecified position at the start
of sectioning in manual input
mode.
・Check that there is no
block in the position
where not specified by
LK PC.
9002
<Error Message>
Unit:
Microcomputer board
Part:
Microcomputer board
Symptom:
Timer abnormality for
remote communication
The timer for remote
communication is not
responding.
・Contact DNS for
service.
9003
<Error Message>
Unit:
LK-400
Part:
LK PC
Symptom:
Remote communication
connection error
The PC cannot communicate
with LK-400.
・Confirm the LK-400
software on the PC is
running.
9010
<Error Message>
Unit:
LK-400
Part:
LK PC
Symptom:
Sample number
receiving error
Issues with the sample number
received by LK PC.
・Confirm that the
software is running on
the PC.
・Confirm that the
communication cable
between the system and
PC is connected.
9011
<Error Message>
Unit:
LK-400
Part:
LK PC
Symptom:
communication error
PST command transmission error
9012
PST command response timeout
9013
Communication disconnection when
PST command is received
9014
PST command response receiving error
9020
PBR command transmission error
9021
PBR command response timeout
9022
Communication disconnection when
PBR command is received
9023
PBR command response receiving
error
9030
PBI command transmission error
9031
PBI command response timeout
9032
PBI command response sample
number error
9033
PBI command response sectioning
condition number error
9034
<Error Message>
Unit: Slide Transfer Unit
Part: Slide feeder
Symptom:
No slide glass
No slide glasses specified by LK PC is
set on the system.
・Check the setting of the
slide glass.